close
Profil Hiduplah seolah engkau mati besok, Belajarlah seolah engkau hidup selamanya.
envato behance dribbble twitter google instagram

Total Organic Carbon Analysis for Cleaning Validation in Pharmaceutical Manufacturing

Posted By Anonim on Jumat, 24 Juli 2015 | Juli 24, 2015



Total Organic Carbon Analysis for Cleaning Validation in Pharmaceutical Manufacturing


Introduction
In the pharmaceutical industry, Good Manufacturing Practices (GMPs) require that the cleaning of drug manufacturing equipment be validated.1 Many different validation techniques can demonstrate that the manufacturing equipment is cleaned and essentially free from residual active drug substances and all cleaning agents.
Common analytical techniques in the validation process include HPLC, spectrophotometry (UV/Vis), and TOC. HPLC and UV/Vis are classified as specific methods that identify and measure appropriate active and substances. TOC is classified as a non-specific method and is ideal for detecting all carbon-containing compounds including active species, excipients, and cleaning agent(s). 2,3,4,5
The disadvantage of specific methods, particularly HLPC, is that a new procedure must be developed for every active drug substance that is manufactured. This development process can be very time consuming and tedious, plus important sampling issues also must be considered. In addition, HPLC analyses must be performed in a relatively short time period after sampling to avoid any chemical deterioration of the active substance. Finally, the sensitivity of HPLC methods can be limited by the presence of degradation products. Of course, the disadvantage to non-specific methods like TOC is that they can not identify exactly what the residue material is. Depending upon the chosen cleaning process and established acceptance limits, a non-specific method may be all that is needed to validate the process.
TOC analysis can be adapted to any drug compound or cleaning agent that contains carbon and is “adequately” soluble in water. Studies have shown that TOC methods also can be applied to carbon containing compounds that have limited water solubility, and recovery results are equal to those achieved by HPLC.6 TOC methods are sensitive to the ppb range and are less time consuming than HPLC or UV/Vis. USP TOC methods are standard for Water for Injection and Purified Water, 7 and simple modifications of these methods can be used for cleaning validation.



Methodology
TOC analysis involves the oxidation of carbon and the detection of the resulting carbon dioxide. A number of different oxidation techniques exist including photocatalytic oxidation, chemical oxidation, and high-temperature combustion. In this study, a TOC Analyzer, equipped with an autosampler, was used. The Analyzer measures TOC in accordance with ASTM methods D 4779-88 and D 4839-88. It measures TOC directly by adding phosphoric acid to the water sample to reduce the pH to approximately 2 to 3. At this low pH, any inorganic carbon that is present is liberated as CO2 into a nitrogen carrier gas and is directly measured by a Non-Dispersive Infra Red (NDIR) detector. Any remaining carbon in the sample is assumed to be TOC. A sodium persulfate oxidant is then added to the sample, and in the presence of UV radiation, the remaining carbon is oxidized to CO2. The amount of CO2 generated is then measured by the NDIR to determine the amount of TOC originally present in the water.
For equipment cleaning validation there are two types of TOC sampling techniques. One is the direct surface sampling of the equipment using a swab. The second consists of a final rinse of the equipment with high-purity water (typically <500 ppb TOC) and collecting a sample of the rinse for analysis. In general, direct surface sampling indicates how clean the actual surface is. The purpose of this study is to demonstrate how to develop and validate a TOC method to measure a variety of different organic residues on stainless steel surfaces. Performance parameters tested include linearity, Method Detection Limit (MDL), limit of Quantitation (LOQ), accuracy, precision, and swab recovery.
Linearity
TOC analysis should provide a linear relationship between the measured compound concentration and the TOC response of the analyzer. We evaluated four different types of cleaning agents for linearity:
CIP-100Ò (alkaline), CIP-200Ò (acidic), AlconoxÒ (emulsifier), and Triton-X 100 (wetting agent). Results are shown in Figures 1-4. Correlation coefficients ranged from 0.9787 to 0.9998. Alconox and Triton-X 100 have a tendency to “foam” depending upon the concentrations that are analyzed, and this “foaming” phenomena can have a negative effect on the accuracy of the TOC result (reduced R2). Three representative examples of active substances also were tested for linearity: an excipient (sucrose), an antibiotic (vancomycin), and endotoxin. Results are shown in Figures 5-7. All three compounds demonstrated excellent linearity with correlation coefficients (R2) ranging from 0.9996 to 0.9998.
 

Blog, Updated at: Juli 24, 2015

0 komentar:

Posting Komentar

Popular Posts